Sensoray 526 Manuale Utente Pagina 14

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 29
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 13
14
=
0
0
CC
CC
VV
ref
meas
refmeas
,
where
ref
V
- the actual value of the on-board 10V reference (from the EEPROM);
meas
C
- ADC reading corresponding to the measured voltage;
ref
C
- ADC reading corresponding to the on-board +10V reference;
0
C
- ADC reading corresponding to the on-board 0V reference.
The calibration constant is stored in the
double
floating point format (8 bytes per value). See
the
Calibration EEPROM
section for the details.